25-OH Vitamin D3 and 25-OH Vitamin D2 were determined by rapid, simple and validated LC/MS/MS method without using SPE and isotopic standard. The relative standard deviations (RSD) were found to be 3.1 and 2.8 % for 25-OH Vitamin D3 and 25-OH Vitamin D2, respectively. The correlation coefficient was grater than 0.99 for both analytes in the calibration range. The recoveries at 100 μg L-1 concentration averaged to 92 % for 25-OH vitamin D3 and 94 % for 25-OH-vitamin D2. The uncertainty of the measurements for 25-OH vitamin D3 and for 25-OH-vitamin D2 was 7.77 % and 7.24 %, respectively.
KEYWORDS- 25-OH Vitamin D3
- 25-OH Vitamin D2
- LC/MS/MS
- method validation
- uncertainty